Tag Archives: reflectometer

testing detectors…

The ISIS detector group has visited our facility to perform some tests with us on their new scintillator detectors. Scintillator detectors are usually more gamma sensitive and less neutron sensitive than the standard 3-He detectors. But since the enormous increase in the price of 3-He, more efforts have been put into the development of scintillator based detectors. The ISIS team came to the RID for tests and demonstration of clear-fibre and wavelength-shifting fibre detectors and a whole set of electronics to evaluate the output of these detectors.

During a three-day test on our reflectometer, we have compared these detectors with the standard 3-He detector of the reflectometer. The preliminary results of these tests show that the neutron detection efficiency is some 70-80% of this 3-He detector and the gamma sensitivity (without optimization of the pulse analysis of the scintillator electronics) is between 20 and 50% higher than the 3-He tube. More involved tests and optimizations need to be done for the gamma-sensitivity. A more severe testing would be counting neutrons in a strong(er) gamma background.

 

detector tests setup

The photo above shows the installation of one of the detectors (the aluminium box) inside the sample-chamber of our neutron-reflectometer at RID. The neutron beam impinges from the left on the scintillators inside the box (not visible) and the light generated by these scintillator crystals is amplified by the photon-multiplier-tubes (PMTs) which are read out by the electronics (not visible). The pulse-shape- and coincidence analysis of these electronics determine whether a neutron or a gamma is detected. The detectors were tested on the reflectometer, so that we could use the time-of-flight option and diafragms of this instrument.